Masaq Index
arXiv 2006-05-23 DOI 10.1016/j.bios.2006.07.011 0 views

Giant magnetoimpedance: new electrochemical option to monitor surface effects?

Kurlyandskaya, G. V. · Miyar, V. Fal

Original · EN

Magnetoimpedance, MI, change due to surface modification of the sensitive element caused by biofluids was studied with the aim of creating a robust sensor capable of separating the chemical surface modification from the sensing process. A MI sensor prototype with an as-quenched FeCoSiB amorphous ribbon sensitive element was designed and calibrated for a frequency range of 0.5 to 10 MHz at an intensity of the current of 60 mA. Measurements as a function of the exposure time were made, first, in a regime where chemical surface modification and sensing were separated and then, in a regime where they were not separated (in a bath for fluids). The MI variation was explained by the change of the surface magnetic anisotropy. It was shown that the magnetoimpedance effect can be successfully employed as a new electrochemical option to probe the electric features of surface-modified magnetic electrodes when the biofluid, the material of the sensitive element, and the detection conditions are properly selected and synergetically adjusted.

English translation

This paper has no Arabic translation yet. Be the first: it takes a few seconds, and the result is stored for every future reader.