Masaq Index
arXiv 2014-11-19 DOI 10.1063/1.4902131 0 views

Dry-Mass Sensing for Microfluidics

Müller, T. · White, D. A. · Knowles, T. P. J.

Original · EN

We present an approach for interfacing an electromechanical sensor with a microfluidic device for the accurate quantification of the dry mass of analytes within microchannels. We show that depositing solutes onto the active surface of a quartz crystal microbalance by means of an on-chip microfluidic spray nozzle and subsequent solvent removal provides the basis for the real-time determination of dry solute mass. Moreover, this detection scheme does not suffer from the decrease in the sensor quality factor and the viscous drag present if the measurement is performed in a liquid environment, yet allows solutions to be analysed. We demonstrate the sensitivity and reliability of our approach by controlled deposition of nanogram levels of salt and protein from a micrometer-sized channel.

English translation

This paper has no Arabic translation yet. Be the first: it takes a few seconds, and the result is stored for every future reader.