An All-Cryogenic THz Transmission Spectrometer
Burke, P. J. · Eisenstein, J. P. · Pfeiffer, L. N. · West, K. W.
الأصل · EN
This paper describes a THz transmission spectrometer for the spectral range of 2-65 cm-1 (100 GHz to 2 THz) with a spectral resolution of at least 1.8 cm-1 (50 GHz) where the source, sample, and detector are all fully contained in a cryogenic environment. Cyclotron emission from a two-dimensional electron gas heated with an electrical current serves as a magnetic field tunable source. The spectrometer is demonstrated at 4.2 K by measuring the resonant cyclotron absorption of a second two dimensional electron gas. Unique aspects of the spectrometer are that 1) an ultra-broadband detector is used and 2) the emitter is run quasi-continuously with a chopping frequency of only 1 Hz. Since optical coupling to room temperature components is not necessary, this technique is compatible with ultra-low temperature (sub 100 mK) operation.