Characterization of a commercial, front-illuminated interline transfer CCD camera for use as a guide camera on a balloon-borne telescope

Clark, Paul · Massey, Richard · Chang, Herrick L. · Galloway, Mathew · Israel, Holger · Jones, Laura L. · Li, Lun · Mandic, Milan · Morris, Tim · Netterfield, Barth · Peacock, John · Sharples, Ray · Susca, Sara

الأصل · EN

We report results obtained during the characterization of a commercial front-illuminated progressive scan interline transfer CCD camera. We demonstrate that the unmodified camera operates successfully in temperature and pressure conditions (-40C, 4mBar) representative of a high altitude balloon mission. We further demonstrate that the centroid of a well-sampled star can be determined to better than 2% of a pixel, even though the CCD is equipped with a microlens array. This device has been selected for use in a closed-loop star-guiding and tip-tilt correction system in the BIT-STABLE balloon mission.