Extreme ultraviolet spectrometer based on a transmission electron microscopy grid
Sistrunk, Emily · Gühr, Markus
Original · EN
We performed extreme ultraviolet spectroscopy using an 80 lines/mm TEM mesh as dispersive element. We present the usefulness of this instrument for dispersing a high harmonic spectrum from the 13th to the 29th harmonic of a Ti:Sapph laser, corresponding to a wavelength range from 60 to 27 nm. The resolution of the instrument is limited by the image size of the high harmonic generation region on the detector. The best resolution in first order diffraction is under 2 nm over the entire spectral range.
English translation
This paper has no Arabic translation yet. Be the first: it takes a few seconds, and the result is stored for every future reader.