SPI energy calibration
Lonjou, V. · Knodlseder, J. · Roques, J. P. · Skinner, G. K. · von Ballmoos, P. · Jean, P. · Paul, P. · Weidenspointner, G. · Wunderer, C. · Schanne, S.
Original · EN
The status of the SPI energy calibration after the first year of INTEGRAL operations is reported. We have studied the gain variations and we have demonstrated that the most important parameter is the germanium detector temperature. This study permits us to determine the limits of our calibration method and the frequency of calibrations needed.
English translation
This paper has no Arabic translation yet. Be the first: it takes a few seconds, and the result is stored for every future reader.