Investigation of background in large-area neutron detectors due to alpha emission from impurities in aluminium

Birch, J. · Buffet, J. -C. · Clergeau, J. -F. · van Esch, P. · Ferraton, M. · Guerard, B. · Hall-Wilton, R. · Hultman, L. · Höglund, C. · Jensen, J. · Khaplanov, A. · Piscitelli, F.

Original · EN

Thermal neutron detector based on films of ¹⁰B₄C have been developed as an alternative to ³He detectors. In particular, The Multi-Grid detector concept is considered for future large area detectors for ESS and ILL instruments. An excellent signal-to-background ratio is essential to attain expected scientific results. Aluminium is the most natural material for the mechanical structure of of the Multi-Grid detector and other similar concepts due to its mechanical and neutronic properties. Due to natural concentration of α emitters, however, the background from α particles misidentified as neutrons can be unacceptably high. We present our experience operating a detector prototype affected by this issue. Monte Carlo simulations have been used to confirm the background as α particles. The issues have been addressed in the more recent implementations of the Multi-Grid detector by the use of purified aluminium as well as Ni-plating of standard aluminium. The result is the reduction in background by two orders of magnitude. A new large-area prototype has been built incorporating these modifications.