A First Comparison of the responses of a He4-based fast-neutron detector and a NE-213 liquid-scintillator reference detector
Jebali, R. · Scherzinger, J. · Annand, J. R. M. · Chandra, R. · Davatz, G. · Fissum, K. G. · Friederich, H. · Gendotti, U. · Hall-Wilton, R. · Håkansson, E. · Kanaki, K. · Lundin, M. · Murer, D. · Nilsson, B. · Rosborg, A. · Svensson, H.
الأصل · EN
A first comparison has been made between the pulse-shape discrimination characteristics of a novel ⁴He-based pressurized scintillation detector and a NE-213 liquid-scintillator reference detector using an Am/Be mixed-field neutron and gamma-ray source and a high-resolution scintillation-pulse digitizer. In particular, the capabilities of the two fast neutron detectors to discriminate between neutrons and gamma-rays were investigated. The NE-213 liquid-scintillator reference cell produced a wide range of scintillation-light yields in response to the gamma-ray field of the source. In stark contrast, due to the size and pressure of the ⁴He gas volume, the ⁴He-based detector registered a maximum scintillation-light yield of 750 keVee to the same gamma-ray field. Pulse-shape discrimination for particles with scintillation-light yields of more than 750 keVee was excellent in the case of the ⁴He-based detector. Above 750 keVee its signal was unambiguously neutron, enabling particle identification based entirely upon the amount of scintillation light produced.