Linear Paul trap design for an optical clock with Coulomb crystals

Herschbach, Norbert · Pyka, Karsten · Keller, Jonas · Mehlstäubler, Tanja E.

Original · EN

We report on the design of a segmented linear Paul trap for optical clock applications using trapped ion Coulomb crystals. For an optical clock with an improved short-term stability and a fractional frequency uncertainty of 10-18, we propose 115In+ ions sympathetically cooled by 172Yb+. We discuss the systematic frequency shifts of such a frequency standard. In particular, we elaborate on high precision calculations of the electric radiofrequency field of the ion trap using the finite element method. These calculations are used to find a scalable design with minimized excess micromotion of the ions at a level at which the corresponding second- order Doppler shift contributes less than 10-18 to the relative uncertainty of the frequency standard.