Optical microscope and tapered fiber coupling apparatus for a dilution refrigerator
MacDonald, A. J. R. · Popowich, G. G. · Hauer, B. D. · Kim, P. H. · Fredrick, A. · Rojas, X. · Doolin, P. · Davis, J. P.
الأصل · EN
We have developed a system for tapered fiber measurements of optomechanical resonators inside a dilution refrigerator, which is compatible with both on- and off-chip devices. Our apparatus features full three-dimensional control of the taper-resonator coupling conditions enabling critical coupling, with an overall fiber transmission efficiency of up to 70%. Notably, our design incorporates an optical microscope system consisting of a coherent bundle of 37,000 optical fibers for real-time imaging of the experiment at a resolution of 1 μm. We present cryogenic optical and optomechanical measurements of resonators coupled to tapered fibers at temperatures as low as 9 mK.
الترجمة العربية
لا توجد ترجمة عربية لهذا البحث بعد. كن أوّل من يطلبها: تستغرق ثوانيَ معدودة، وتُحفظ النتيجة لكل قارئ قادم.