Construction and measurements of a vacuum-swing-adsorption radon-mitigation system
Schnee, R. W. · Bunker, R. · Ghulam, G. · Jardin, D. · Kos, M. · Tenney, A. S.
الأصل · EN
Long-lived alpha and beta emitters in the ²²²Rn decay chain on (and near) detector surfaces may be the limiting background in many experiments attempting to detect dark matter or neutrinoless double-beta decay, and in screening detectors. In order to reduce backgrounds from radon-daughter plate-out onto the wires of the BetaCage during its assembly, an ultra-low-radon cleanroom is being commissioned at Syracuse University using a vacuum-swing-adsorption radon-mitigation system. The radon filter shows 20× reduction at its output, from 7.47±0.56 to 0.37±0.12 Bq/m³, and the cleanroom radon activity meets project requirements, with a lowest achieved value consistent with that of the filter, and levels consistently < 2 Bq/m³.
الترجمة العربية
لا توجد ترجمة عربية لهذا البحث بعد. كن أوّل من يطلبها: تستغرق ثوانيَ معدودة، وتُحفظ النتيجة لكل قارئ قادم.